Specification | getProbe DR LS |
Light source / Optics |
Measurement geometry | Diiffuse reflectance |
Light source | 2 redundant tungsten light sources
0,5 - 4,5W, programmable |
Lamp life time (each) | 18 months - 5 years
(depending on operating conditions) |
Wavelength range | 400-2500nm |
Spot diameter | 6 mm at sample window |
Internal reference | PTFE based standard + wavelength reference |
Detection fibre | SMA-port for single 400 or 600µm fibres, up to 100m |
Probe electronics |
Power supply | 8 - 24 VDC, < 8W (standard configuration) |
Control interface | EIA-485 (RS-485) |
Lamp regulation | Power, voltage or current controlled, soft start |
Additional functions | Lamp selection, reference switch control, probe temperature sensor, EEPROM stored ratio to external reference, operation hours counter |
Process conditions |
Probe tip temperature | 0°C - 180°C |
Environmental conditions | 0°C - 60°C |
Pressure range | 0 - 40 Bar |
Process connections | DIN- or ANSI flanges welded on the probe
Swagelok, other standard fittings |
Materials and dimensions |
Probe body material | Stainless steel 1.4404 / AISI 315L, polished Ra<0.2 |
Probe housing material | Aluminium anodized |
Window material | Sapphire |
Dimensions (mm) |