HR4000 - Wavelength Ranges & Grating Selection



Grating Selection Chart

The following gratings are available:

Grating #ApplicationLines/mmBlaze WavelengthMax. Efficiency (>30%)
H1UV 600300 nm200-575 nm
H2UV/VIS600400 nm250-800 nm
H3VIS/Color600500 nm350-850 nm
H4NIR600750 nm530-1100 nm
H5UV/VIS1200Holographic UV200-400 nm
H6NIR1200750 nm500-1100 nm
H7UV/VIS2400Holographic UV200-500 nm
H9VIS/NIR1200Holographic VIS400-800 nm
H10UV/VIS1800Holographic UV200-635 nm
H11UV/VIS1800Holographic VIS320-800 nm
H12UV/VIS2400Holographic VIS200-575 nm
H13UV/VIS/NIR300500 nm300-1100 nm
H14NIR 6001000 nm650-1100 nm



Example - 600 lpm Grating, 5 µm Slit:
If your starting wavelength is at 400 nm, then your spectral range is ~437 nm, which gives you a 400-837 nm wavelength range and 0.239 nm resolution.


Example - 1200 lpm Grating, 5 µm Slit:
If your starting wavelength is at 600 nm, then your spectral range is ~188 nm, which gives you a 600-788 nm wavelength range and 0.104 nm resolution.



Example - 1800 lpm Grating, 5 µm Slit:
If your starting wavelength is at 700 nm, then your spectral range is ~90 nm, which gives you a 700-790 nm wavelength range and 0.050 nm resolution.



Example - 2400 lpm Grating:
If your starting wavelength is at 500 nm, then your spectral range is ~71 nm, which gives you a 500-571 nm wavelength range and 0.038 nm resolution.